C/N:ES 106


1.BACKGROUND

iCheck®-DON strip is lateral flow immunochromatographic assay based on a competitive immunoassay format. With iCheck®strip reader, it measures 500-5000μg/kg(ppb) of Deoxynivalenol in corn or rice.


2.MATERIALS PROVIDED

1) iCheck®-DON strip (25/pk).
2) Microwells coated (25)
3) DON Dilution buffer (5ml)
4) 200µl Pipette Tip (50)
5) DON Calibration RFID card
6) Product Manual
Materials required but not provided:
1) 200 and 1000µl precision micropipette
2) iCheck®Strip reader (with ZERO card)
3) Multi-purpose rotary mixer
4) Timer
5) Weigh scale
6) Distilled water ot pure water
7) Funnel
8) Filter paper
9) 50mL Extraction tube


3.PRECAUTIONS
1) Store test kits at 2~8℃(35.6~46.4℉) when not in use, and do not use beyond the expiration date.
2) Keep from moisture, and do not use the damaged and damp kits.
3) All reagents must be at room temperature before assay is running.
4) Do not re-use test strips.
5) Avoid swallowing the desiccant.


4.INTERPRETATION OF THE RESULTS
1)Test range: 500-5000ppb.
2) If there is no line in control zone, the test is invalid and the sample should be re-tested by using a valid test strip.
3) If reading is 5000 ppb, the filtered extract should be further diluted in pure water such that the diluted sample reading results are in a range of 500-5000ppb and reanalyzed to obtain accurate results. The dilution factor must be included when the final result is calculated.


INQUIRY

If you have any suggestions or opinions about our products, please leave a message, and we will immediately answer your questions. Thanks for your support.

  • *
    *
  • *
  • *
Name cannot be empty!
Please enter a valid phone number!
Please enter a valid E-mail!
The message content cannot be empty!
Submitted Successfully
Submission Failed

Copyright ©Clover Technology Group, Inc. 2025All Rights Reserved